![]() |
ENVIRONMENTAL
EDUCATION SERIES
ENVIRONMENTAL QUALITY Agriculture & Natural Resources |
|
|
|
The pounds of material applied per acre for large volumes (10 pounds and up) can be determined by using the following procedure. This calibration procedure is based on 1/16 of an acre, which is equal to 16 ounces in a pound of material. The measured distances shown in Table 2 correspond to those found in liquid sprayer calibration procedures, but the material collected is multiplied by 8 to obtain rate per acre.
A weight scale incremented in ounces is needed
for this procedure. Check uniformity of outlets across the swath or row.
Collect from each outlet for a specific time and weigh the material. Each
outlet should be within 5 percent of the average output. Extreme care should
be exercised and protective equipment used when an active ingredient is
involved.
| Step 1. | Determine the type of application to be made and select appropriate procedure from Table 1. Example: Broadcast - Procedure A |
Table 1.
| Type of
Fertilizer Application |
Procedure | Coverage Basis
(volume of application) |
|
|
||
| Broadcast
Band |
A
B |
Broadcast (lbs./acre)
Broadcast (Lbs./acre of band) |
|
|
||
| Note: Determine and use average row spacing for modified row patterns. Use width of area covered per row spacing in skip row patterns for broadcast rates. | ||
| Step 2. | Using procedure A or B as selected in Step 1, determine appropriate calibration distance from Table 2. |
| (A) | Broadcast Application: Outlets must be evenly spaced. Measure outlet spacing. Find this spacing in left column of Table 2, and read the corresponding calibration distance. Example: For a 19-inch spacing, the distance would be 215 feet. |
| (B) | Band Application: Measure band width. Find this band width in the left column of Table 2, and read the corresponding calibration distance. Example: For a 12 inch band, the distance would be 340-feet. |
Table 2.
|
|
|
| Row Spacing or Outlet Spacing
or Band Width (whichever applies) |
Calibration
Distance |
|
|
|
| 48 inches
46 44 42 40 38 36 34 32 30 24 20 19 18 12 10 8 |
85 feet
89 93 97 102 108 113 120 128 136 170 204 215 227 340 408 511 |
|
|
|
Example: For a 13-inch band, the calibration distance would be 340 divided by 13/12 = 314 feet.Note: To increase calibration accuracy for a wide outlet spacing, multiply calibration distance by a factor (for example, 2). Then, divide Step 7 by the same factor for pounds per acre. For narrow spacings with long calibration distances, divide calibration distance by a factor (for example, 4). Then, multiply Step 7 by the same factor for pounds per acre.
| Step 3. | Measure and mark calibration distance in a typical field application site. |
| Step 4. | With all attachments in operation (harrow, planters, etc.) and traveling at the desired operating speed, determine the number of seconds it takes to travel the calibration distance. Be sure machinery is traveling at full operation speed the full length of the calibration distance. Mark or make note of engine RPM. Machine must be operating at same speed used for calibration. |
| Step 5. | With applicator sitting still and operating at same speed as used in Step 4, adjust gate openings to desired setting. |
| Step 6a. | For procedure (A), Step 2, broadcast application, collect from one outlet for the number of seconds required to travel the calibration distance. For procedure (B), Step 2, band application, collect from all outlets used on one band width for the number of seconds required to travel the calibration distance. |
| Step 7. | Weigh the amount of materials collected in ounces. The number of ounces collected multiplied by 8 is the pounds per acre rate on the coverage basis, as indicated in Table 1. For example, if you collect 40 ounces using procedure (A) or (B), the applicator will apply 40 multiplied by 8 = 320 pounds per acre on a broadcast coverage basis. Adjust applicator speed, gate opening, etc.. to obtain recommended rate. |
| Step 8. | Check applicator for proper calibration every four to eight hours of use. Simply repeat steps 6 and 7. If there is a difference of more than 5 percent of original calibration, check the system. |
Band Application
To determine the pounds of material required to make a band application
on a field, the number of acres that will be in the actually treated band
must be determined. When all treated bands are the same width and all untreated
bands are the same width, which is usually the case, the acres in the actually
treated band can be calculated by placing the width of the treated band
over the sum of the treated and untreated band, and multiplying this fraction
by the number of acres in the field. Example: How many acres will actually
be treated in a 30 acre field if a 12-inch band of material is applied
over the drill of rows spaced 36 inches apart? The treated band width
is 12 inches. The untreated band width is 24 inches. Acres actually treated
will be:
| 12"
12" + 24" |
x 30 acres = 10 acres |
The amount of material required will be 10 times the number of pounds per acre from Step 7.
Carefully check rate recommendations as to type of application, broadcast,
band, and type of material specified (formulated product, active ingredients,
etc.).